About this trial
Portal hypertension is characterised by an increased portal pressure gradient (PPG), that is the difference in pressure between the portal vein and the inferior vena cava (IVC). Portal hypertension is a consequence of cirrhosis resulting from chronic hepatitis. Patients with portal hypertension are at risk of developing complications including oesophageal or gastric varices, variceal bleeding, ascites, spontaneous bacterial peritonitis, hepatic encephalopathy and mortality.
Albeit its clinical significance, direct measurement of portal venous pressure to document portal hypertension has traditionally been difficult. The portal vein pressure can be measured by transhepatic or transvenous methods but the procedure carries a risk of intra-peritoneal bleeding. Furthermore, the IVC pressure measurement requires further transjugular catheterisation. Hence, the technique is rarely used. Currently, the gold standard in measurement of portal hypertension is via measurement hepatic venous pressure gradient (HVPG). The HVPG has been shown to correlate with risk of clinical decompensation, development of varices, hepatocellular carcinoma, variceal bleeding, spontaneous bacterial peritonitis and mortality. Nevertheless, the technique has a low acceptance rate amongst patients and it may not be available even in tertiary medical centres.
Recently, the use of EUS-guided approach for measurement of portal pressure gradient (PPGM) has been shown to be feasible. The technical success rate was 100% and no adverse events were reported. Measurements obtained with the EUS approach was shown to correlate excellently with clinical parameters of portal hypertension including presence of varices, portal hypertensive gastropathy and thrombocytopenia. Furthermore, the procedure could be performed at the same time of screening oesophagogastroduodenoscopy (OGD), that is frequently required for variceal screening in this group of patients. Hence, the aim of the current study is to investigate the feasibility of EUS-PPGM and correlate the risk of developing complications with the PPGM in patients that are suffering from chronic hepatitis.
Eligibility criteria
This trial does not accept healthy volunteersQualifiers
Suffering from chronic hepatitis induced cirrhosis
Informed consent available
Disqualifiers
Medical
Child-Pugh Class C
Uncorrected platelet count <50,000
INR > 1.5 (natural)
Trial design
Single group
Treatments tested in this trial
EUS-portal pressure gradient measurement (PPGM)
DeviceEUS-guided PPGM is performed under conscious sedation. An OGD would be performed to evaluate for findings consistent with portal hypertension such as varices and portal hypertensive gastropathy. The linear array echoendoscope is passed from the mouth into the stomach and duodenum, as with a standard exam. Under endoscopic ultrasound guidance, a 25 gauge EUS needle (Echotip, Cook Medical, USA) primed with heparin 100 USP/ml is then placed across the stomach or duodenal wall and through the liver parenchyma into the portal vein. A pressure measurement is then obtained from the compact pressure transducer attached to the handle of the needle. Three separate measurements are made with re-priming of heparin (less than 0.5 ml). The process is repeated for the hepatic vein. The EUS - PPG is calculated by subtracting the average of the three hepatic vein pressure measurements from the average of the 3 direct portal vein pressure measurements.
Treatment groups
Trial outcomes
Primary outcomes
Technical success
Ability to place a needle in the portal and hepatic vein for calculation of portal pressure gradient
Secondary outcomes
Adverse events
Any untoward event directly arising from the measurement of PPGM