Validating Dozee for Contactless Monitoring of Heart, Breathing, and Movemen

Trial statusRecruiting
Trial phaseNot listed
Trial typeObservational
Biological sexAll
Age18-100
SponsorTurtle Shell Technologies Pvt. Ltd.

About this trial

The Dozee VS system (K221555) is a contactless device that tracks heart rate and breathing rate using a sensor sheet placed under the mattress. It captures body vibrations and converts them into vital signs, which can be monitored remotely in real time. The system also supports add-ons like a blood pressure cuff, pulse oximeter, and thermometer, offering a complete monitoring solution. This study aims to validate the updated AI-powered heart rate and improved breathing rate algorithms across different sensor sheet types to confirm clinical accuracy

Eligibility criteria

This trial accepts healthy volunteers

Qualifiers

Adults Male or female. (Above 18 years of age)

Provide written informed consent.

Weight between 40kg/88.18lbs and 120kg/264.55lbs

Agree to not eat during the testing period

Disqualifiers

Are connected to a device that may interfere with the device monitoring in this study.

Are receiving any bedside care which may be incompatible with the study procedures.

A likely need to receive or undergo a procedure during the testing period.

Cannot accept a nasal cannula, or have an ECG leads placed on the chest.

Trial population

The study will be conducted on atleast 60 subjects in the age group above 18 years regardless of gender. The patients/volunteers be informed about the study and a written informed consent form will be obtained from them. Dozee will be placed under the mattress of the patient's bed and the data will be collected through the device whenever a patient is on the bed.

Trial design

Design model

Cohort

Time perspective

Prospective

Treatments tested in this trial

  • Device: Telemetry monitoring

    Device

    All participants will be on standard telemetry monitoring for Heart Rate, Respiration Rate and movements using the gold standard (Electrocardiography and Capnography respectively).

Treatment groups

60 Participants
are divided into 1 treatment group
Group A: Single Arm Study (Cohort)1 intervention

Trial outcomes

Primary outcomes

1

Mean Absolute Error (MAE) of Dozee's Heart Rate a from the Gold Standard ECG

Heart Rate collected at 1 Min intervals from gold standard and Dozee. Mean Absolute Error (MAE) of the Dozee Heart Rate will be calculated.

Time frame
6 months
2

Accuracy (%) of Dozee in identfying subject movement

Time stamps of subject movement during the study will be recorded manullay. Time stamps of Dozee's movement will be recorded. Accuracy of the Dozee device in detecting movement compared to manual observation will be computed in terms of percentage.

Time frame
6 months
3

Accuracy(%) of of Dozee's bed-exits

Time stamps of bed-exits will be recorded manually during the study will be recorded. Time stamps of Dozee's bed-exit will be computed. Accuracy of the Dozee device in detecting bed-exit compared to manual observation as gold standard will be computed in terms of percentage.

Time frame
6 months
4

Mean Absolute Error (MAE) of Dozee's Respiratory Rate from the Gold Standard Capnography

Respiratory Rate collected at 1 Min intervals from gold standard and Dozee. Mean Absolute Error (MAE) of the Dozee Respiratory Rate will be calculated.

Time frame
6 Months

Secondary outcomes

1

Range of values (minimum, maximum, and spread) of Heart Rate in beats per min measured by the gold standard and the Dozee

The outcome will be assessed by comparing the range of values (minimum, maximum, and spread) of Heart Rate ( beats per minute) measured by the gold standard and the Dozee over the study period.

Time frame
6 months
2

Range of values (minimum, maximum, and spread) of Respiratory Rate in breaths per min measured by the gold standard and the Dozee

The outcome will be assessed by comparing the range of values (minimum, maximum, and spread) of Respiratory Rate ( breaths per minute) measured by the gold standard and the Dozee over the study period.

Time frame
6 Months
3

Bland- Altman Level of Agreement between Dozee Heart Rate and Respiratory Rate with the Gold Standard

Level of Agreement between Dozee's HR and RR with reference device computed by Bland Altman.

Time frame
6 months
4

Detection rate (%) of the Dozee's Heart Rate and Respiratory Rate

The time-stamps of Dozee's Heart Rate and Respiratory Rate and the gold standard Heart Rate and Respiratory Rate will be recorded. DR is the proportion of gold standard points for which the DS successfully provides a corresponding measured value, relative to the total number of gold standard points.

Time frame
6 months

Other outcomes

Sponsors and contacts

Click on the lead sponsor to view all of their trials.

Turtle Shell Technologies Pvt. Ltd.

Lead sponsor

Eastside Research Associates

Collaborator